11.0 Spécifications Techniques - Ametek JOFRA ASC-400 Manuel D'utilisation

Calibrateur multifonctions
Table des Matières

Publicité

11.0 Spécifications techniques
Toutes les spécifications sont données pour une température comprises entre 23 ±5°C / 73
±9°F
Mechanical specifications
Description
Operating temperature :
Storage temperature :
Dimension LxWxH :
Case protection:
Humidity:
Display:
Weight incl. Batteries :
Unit in soft case :
Weight incl. Test leas and test chips :
Shipping size :
Shipping weight :
Electrical specifications
Description
Mains adapter:
Batteries :
Battery lifetime, Alkaline, backlight low:
Battery lifetime, Alkaline, backlight high,
12 mA: loop power:
Battery charge current. Use only NiMH cells
with capacity larger than 1700 mAh.:
RS232 communication interface :
Connector :
Communication type :
Display update rate:
Temperature coefficient:
-10°C to 18°C and 28°C to 50°C:
Thermocouple mV
TC mV read
TC mV source
Maximum output current is 3 mA Output impedance 0.010 ohm. Input impedance 10 Mohm.
128797 02
Value
-10 to 50°C / 14 to 122°F
-20 to 60°C / -4 to 140°F
220x96x55 mm / 8.66x3.78x2.17 in
IP40
0% to 80% R.H. non-condensing
2.8" 320*240 TFT colour display
584 g / 20.6 oz
235x95x115 mm / 9.25x3.74x4.53 in
933 g / 32.91 oz
275x100x175 mm / 10.83x3.94x6.89 in
1233 g /43.49 oz
Value
(option) 9VDC/500mA - 230VAC/115VAC
6 x AA batteries
1.5V AA... Alkaline (non rechargeable) or AA NiMh
(rechargeable)
30 hours
13 hours
85 mA
Mini USB female (B)
USB 2.0 / ASCII
2.5/second (mA, V, resistance, RTD, TC)
10/second (pressure, switch-test)
0.003%FS/°C / 0.0017%FS/°F (mA, V, resistance)
0.001%rdg/°C / 0.0006%rdg/°F (frequency)
Range
Min
Max
-10.000 mV
75.000 mV
-10.000 mV
75.000 mV
2014-08-22
Accuracy ±
12 months
0.015% rdg +10µV
0.015% rdg +10µV
60