Thermocouple Cold junction
CJC compensation
CJC outside above
Volt V
Read (Isolated)
Read (non-isolated)
Source
Maximum output current in voltage ranges is 3 mA Output impedance <1 ohm . Input resistance 1 Mohm.
Frequency Pulse
CPM read
Hz read
KHz read
CPM source
Hz source
KHz source
Pulse (source only)
Rate : 1 Hz to 10KHz
Input voltage amplitude range on frequency is 1 to 20 V, Trigger level 0.2 to 10 volt. Minimum pulse width 10
µS.
Output amplitude is adjustable from 1 to 20 V and is a square wave with a 50% duty cycle.
For output frequency, a slight negative offset of approximately -0.1 V is present to assure zero crossing.
Ohm
Ohm read (low)
Ohm read (high)
Ohm source (low)
@ 0.1 to 0.2 mA
@ 0.2 to 0.5 mA
@ 0.5 to IE max
Ohm source (high)
@ 0.05 to 0.1 mA
@ 0.1 to IE max
True Ohm Measurement current (pulsed) 0.25 mA.
3W measurement current match 1% Source exitation current IEXI(max) = 2.0 V / R, IEXI must never exceed 3
mA.
Pulsed current (source) Unit is compatible with smart transmitters and PLCs with pulse > 5 ms.
128797 02
Range
Min
Max
18°C / 64°F
28°C / 83°F
Range
Min
Max
0.000 V
30.000 V
0.000 V
20.000 V
0.000 V
20.000 V
Range
Min
Max
2.0
600.0
0.050
10.000
10.000
100.00
100.00
1000.0
1000.0
10000
1.000
10.000
2.0
600.0
0.050
1000.0
1000.0
10000
1.000
10.000
1
99999
Range
Min
Max
0.00
400.00
400.0
4000.0
5.0
400.00
5.0
400.00
5.0
400.00
400.0
4000.0
400.0
4000.0
2014-08-22
Accuracy ±
12 months
0.2°C / 0.36°F
0.05°C/°C
0.05°F/°F
Accuracy ±
12 months
0.01% rdg +2mV
0.01% rdg +2mV
0.01% rdg +2mV
Accuracy ±
12 months
0.05% rdg + 0.1CPM
0.05% rdg + 0.001Hz
0.05% rdg + 0.01Hz
0.05% rdg + 0.1Hz
0.05% rdg + 1Hz
0.05% rdg +0.001KHz
0.05% rdg
0.05% rdg
0.06% rdg
0.06% rdg
Accuracy ±
12 months
0.015% rdg + 0.03 ohm
0.015% rdg + 0.3 ohm
0.015% rdg +0.10 ohm
0.015% rdg +0.05 ohm
0.015% rdg +0.03 ohm
0.015% rdg +0.5 ohm
0.015% rdg +0.3 ohm
61