Ametek JOFRA ASC-400 Manuel D'utilisation page 62

Calibrateur multifonctions
Table des Matières

Publicité

Current – mA and loop
Description
Range mA:
Loop power for transmitter:
Isolated input:
Current mA
Read (Isolated)
Read (non-isolated)
Source
Hart resistor 250 ohm (On/Off in software). Maximum loop resistance source @ 24 mA (Hart on/ Hart off) 650
ohm / 900 ohm.
mA source voltage input range (external power/HART resistor off) 1V ‐ 30V.
128797 02
Value
0 to 24 mA
Yes, 24 VDC / ±10%
Yes
Range
Min
Max
0.000 mA
24.000 mA
0.000 mA
24.000 mA
0.000 mA
24.000 mA
2014-08-22
Accuracy ±
12 months
0.010% rdg +2µA
0.010% rdg +2µA
0.010% rdg +2µA