10.5 Starting the sterilization cycle

After the steps listed above, choose the most suitable STERILISATION program for the
prepared load, by pressing the
After choosing the program, start the cycle by pressing the
automatically, and the cycle will begin.
During the cycle the display shows all parameters and information related to the cycle in
progress. In this setup the display shows: the type of cycle, the cycle status, the time
remaining before the end of cycle (for the Vacuum Test it identifies the whole cycle, while
for all the other cycles it identifies the sterilization phase plus that of drying), the number of
cycles done by the machine and the button
parameters.

10.6 End of cycle

An acoustic signal will inform the operators about the finished STERILISATION cycle and
the display will show the END CYCLE icon and message.
Unlock the door by pressing the
three multi-function buttons. Should there be pressure present inside the chamber, the
button will not trigger the unlocking. Wait for the complete depressurization of the chamber
and repeat the operation. With the door unlocked, pull the door handle and open.
10.7 Unloading the sterilized materials
Wear personal protective equipment in compliance with regulations on safety and hygiene
at work. Extract the trays by using the spanner provided (FIG.5/6), let rest the instruments
and store them in environments which are not exposed to contamination.
SELECT CYCLE
button.
Info
Unlock
button displayed on the display by one of the
Start
button. The door will lock
allowing access to the list of working
ENGLISH
19