Safety and Regulatory Information
Transportation and Storage
3.
3
• The Applanation Tonometer conforms to ISO 15004-1:2006 for Storage and
Transport Conditions and should always be stored in its carry case when not
in use.
Transport Conditions
Temperature:
Relative Humidity:
Atmospheric Pressure:
Vibration, Sinusoidal:
Shock:
Bump:
Storage Conditions
Temperature:
Relative Humidity:
Atmospheric Pressure:
• The Applanation Tonometer should be checked for damage after unpacking.
• Defective equipment should always be returned in appropriate packaging.
• The equipment may only be operated with the listed accessories (see page 33)
according to the corresponding operating instructions.
Notes on Usage
3.
4
• Only qualified and trained personnel may operate the equipment.
• The training of the operating personnel is the responsibility of the equipment user.
• This type of sensitive measuring equipment should be checked after exposure
to external force (e.g. unintentional shocks or being allowed to fall) and if
necessary, should be returned to the manufacturer for repair.
• The manufacturers of the equipment will not be liable for any loss or damage
resulting from unauthorized intervention.
In particular, it is forbidden to loosen any screws; this may reduce the accuracy
of the measurement. All ensuing warranty claims will be deemed null and void as
a result of unauthorized intervention.
All manuals and user guides at all-guides.com
-40˚C to +70˚C
10% to 95%
500hPa to 1060hPa
10Hz to 500Hz: 0.5g
30g, duration 6 ms
10g, duration 6 ms
-10˚C to +55˚C
10% t0 95%
700Pha to 1060hPa
en
- continued