Chapter 4 - Technical Specification
4.1.2 Connections
These connections are
NOT
1
M13385001
2
3
FIGURE 18: Sample of the connection label
No. Description
Mains power cable and switch
1
Protective earth must be fitted.
l
U.K. Power Cable Fuse: 13A
l
2
Network cable
3
Hand control and Optos Service use
4
Touch screen
4.1.3 Scan head dimensions
The scan head, hand control and touch screen should be placed on a table which meets the table specification,
see Table requirements on the facing page . There should be adequate clearance around the table to operate
the device.
The image server, switch and any browser-based review clients or monitors must be positioned outside the
patient environment; more than 1.5 meters (5 ft) from the scan head.
TABLE 8: Scan head dimensions
Dimension
Width
Depth
Height
Weight
Door access width
Page 64 of 76
English
intended to be detached by the user.
100-240V AC
T5A H250V
4
Details
550 mm (22 inches)
570 mm (23 inches) including chin rest
608 mm to 632 mm (24 inches to 25 inches)
40 kg (89 lbs)
760 mm (30 inches)