Chapter 1 - Warnings and cautions
1.3 Electrical safety
Equipment classification
Electrical Shock Protection - Class 1 Equipment
Type B - relates to the allowable maximum leakage current which can flow from the
applied part. The limits for this are defined in IEC/EN 60601.
The applied parts are the forehead pad, chin pad and face pad.
Medical device. Continuous operation.
Do
NOT
inside.
Do
NOT
plugged directly into a wall socket.
Do
NOT
To prevent electric shock:
WARNING
l
l
l
l
l
If the mains cable is damaged do not use the system and contact your Optos
representative, see Contact us on page 71.
The image server, switch and any browser-based review clients or monitors must be
positioned outside the patient environment; more than 1.5 meters (5 ft) from the scan
head.
Do
NOT
device. Independently sourced peripherals may
representative for more information on compatible peripherals, see Contact us on page 71.
Ordinary equipment (IP2X), without protection against ingress of water.
You must regularly inspect the equipment. If there is any damage please contact your
Optos representative, see Contact us on page 71.
CAUTION
Ensure the mains socket is easily accessible.
Page 14 of 76
English
open the scan head covers. There are no user serviceable or replaceable parts
plug the scan head into a power block or powered table. The scan head must be
disconnect or connect the touch screen while the scan head is in use.
Only use cables supplied by Optos.
Cables should
be extended or altered.
NOT
Mains supply must be earthed.
The device may only be serviced by Optos trained personnel.
Care must be taken never to touch exposed parts of the device while in physical
contact with a patient.
connect any peripherals to the scan head. Only use peripherals supplied with the
be compatible. Contact your Optos
NOT