Chapter 3 - How to...
3.2.8.2 OCT adjustment controls
You can adjust the center, focus, and location of the captured OCT images.
1
2
3
4
FIGURE 14: OCT adjustment controls
Signal to Noise Ratio
1
SNR
has a major impact on OCT image quality and therefore could determine the ability
CAUTION
to distinguish the retinal layers, see SNR and OCT scan quality on page 49.
No. Description
Live retina view.
1
Before capturing a scan this shows the live SLO feed
l
After capturing a scan this shows the SLO alignment at time of capture
l
Tracking status.
2
If
Tracking
l
If
Tracking
l
1. Signal to noise ratio
Page 46 of 76
English
11
5
is displayed the scan location is consistent with the planned location
is
displayed the scan location is
NOT
6
aligned with the planned location
NOT
Part Number: G108707/8GCA
Copyright 2018, Optos plc. All rights reserved.
10
9
8
7