8.2.
installation on the pipe
danger
Risk of injury due to high pressure in the installation.
•
Stop the circulation of fluid and depressurize the pipes before
loosening the fittings.
Risk of injury due to high fluid temperatures.
•
Use safety gloves to handle the device.
Risk of injury due to the nature of the fluid.
•
Respect the regulations on accident prevention and safety rela-
ting to the use of aggressive fluids.
If the conductivity is measured in liquids containing solids
that may leave deposits in the bottom of the pipe, use
installation position 1 (see Fig. 11).
The neutrino transmitter 8222 is available with an external threaded
sensor or with a G 1 1/2'' fastening nut. The transmitter with a
nut can be installed on a pipe by means of an S022 adapter or a
Ø 32 mm true union fitting. The transmitter with an external threaded
sensor can be installed on a pipe or a tank wall by means of an
internal threaded adapter.
22
All manuals and user guides at all-guides.com
8.2.1.
→
Fig. 11:
→
Type 8222 ELEMENT neutrino
Installationandwiring
installation of a transmitter with a
G 1 1/2" fastening nut
Prefer "1" mounting to install a transmitter with sensor
C=0,1 or C=0,01.
Choose an appropriate position in the pipe to install the fitting
(see Fig. 11).
1
2
Mounting positions of the fitting/transmitter unit in the pipe.
Install the transmitter in the fitting (see
english
direction of the fluid
Fig. 12).