2.
Device acquires the flow signal, transforms it to L/s and sends it via USB or Bluetooth to
the software. When the spirometry maneuver finishes, device also calculates parameters
and adjusts graph. Parameters and graphs are sent to the external software under
request. This mode is only available for FVC tests.
DATOSPIR
is powered with two AAA NiMH or Alcaline batteries or through the USB
port when connected to a computer.
DATOSPIR AIRA is NOT designed for use under other conditions or using other power sources
not indicated in this User's Manual. Only the accessories specified in the manual must be used.
1.3 USER PROFILE
DATOSPIR
has been designed for its use by health professionals, being supervised
or instructed by a physician. Specific training in the spirometry technique is recommended.
The Bronchoconstriction test must be supervised by a technician qualified in this technique.
The user must be familiar with the device functioning before using it on patients. All the
required information for the correct use of the device is available in this User Manual.
Although the patient can handle the device while blowing, the software application must
be managed by a physician or a trained technician in spirometry.
DATOSPIR
is indicated for patients older than 4 years, with weight over 15 Kg and
height over 50 cm, and with a mental and physical condition allowing the performance of
the forced maneuver.
1.4 LIMITATIONS FOR USE. CONTRAINDICATIONS
The analysis of the results of a spirometry test is not enough to make a correct diagnostic
about the clinical condition of the patient. Interpretation of tests must be complemented
by the clinical history or other test that the doctor considers appropriate for deciding the
correct treatment.
The collaboration of the patient is required to make a spirometry test. Complete forced
expiration is necessary to obtain significant FVC values. The physician must assess the
patient's ability to perform these tests. Pay special attention to children, the elderly and
people with disabilities.
Medical staff must consider the symptoms showed by the patient. The acceptability of a
test is the responsibility of the sanitary staff.
511-D00-MUM · Rev 1.02
DATOSPIR
· User's Manual
EN 5