LERBAKKEN 10, 8410 RØNDE DENMARK
Warning: To prevent ignition of the explosive atmospheres, disconnect power before servicing and do not
separate connectors when energized and an explosive gas mixture is present.
Warning: Do not install or remove modules from the Power Rail when an explosive gas mixture is present.
Hazardous Classified Location
Class I/II/III, Division 1, Group A,B,C,D,E,F,G
or Class I, Zone 0/1 Group IIC, [AEx ia] IIC
or Group IIC, [Ex ia Ga] IIC Gc
Simple Apparatus or
Intrinsic safe apparatus
with entity parameters:
Vmax (Ui) ≥ Vt (Uo)
Imax (Ii) ≥ It (Io)
Pi ≥ Pt (Po)
Ca ≥ C
+ Ci
cable
La ≥ L
+ Li
cable
CH1 (terminal 41,42)
CH2 (terminal 51,52)
U
,Voc:
28 V
o
I
, Isc:
93 mA
o
P
:
0.65 W
o
IIC or A,B
C
0.08F
o.
L
4 mH
o.
Revision date:
2019-04-04
28
-20 ≤Ta ≤ +60ºC
+
mA
_
+
mA
_
IIB or C,E,F
IIA or D,G
0.650F
2.15 F
16 mH
32 mH
Version Revision
V3 R0
9107 - Version de produit 9107-002
Unclassified Location or
Hazardous Classified Location
Class I, Division 2, Group A,B,C,D T4
or Class I Zone 2 Group IIC T4 Gc
44
43
42
CH1
41
54
53
52
CH2
51
91 92 93 94 95
Power Rail
Prepared by:
PB
9107QF01
4501
34
33
32
31
14
13
12
11
9107
(terminal 11,12,13,14)
(terminal 31,32,33,34)
(terminal 91,92,93,94,95)
Um: 253 V, max 400 Hz
Page:
2/2