INSTALLATION – OPERATION – MAINTENANCE
OPERATING CHECKS
AND INSPECTION
PACKAGING
STEAM
STERILIZATION
MAXIMUM DURATION
STORAGE
9.8. Replacing the filter in the separation vessel
1
10.

PUTTING OUT OF OPERATION

When the dental aspirator is not used for a longer time, it is necessary to carry out the cleaning and
disinfection of all parts as described in the Articles No. 9.1, 9.3 and then turn the power-supply switch
(5) on in the position "I", take the suction tubings (6) off the holder (2) and let the air intake flow
through (for approximately 15-20 min) so that the suction system is fully dried out. Afterwards put
the tubings into the holder and turn off the power-supply switch in the position "0", disconnect the
equipment from the network (unplug) and remove the suction mouthpieces (1) from suction tubings.
DISPOSAL
11.
DISPOSAL OF APPLIANCE
 Disconnect the appliance from mains.
 To follow the rules of personal hygiene for works with contaminated material
 Clean device according, section 9.
NP-DO M-37_10-2020
All manuals and user guides at all-guides.com
 Follow the manufacturer´s instructions concerning selection of
cleaning/disinfection cycles and detergents to be used.
 On completing the cycle, check there is no dirt residue on the
product; repeat the cleaning cycle whenever necessary.
Visually inspect the condition and cleanliness level of the product.
Discard all damaged or worn products.
Select EC marked packaging that meets the requirements of Council
Directive 93/42/CEE and addenda, suitable for the product and the
sterilization procedure, leaving enough space to prevent tension
when sealed.
Fractionated pre-vacuum steam sterilization using EN 13060 or EN
285 standard compliant equipment and validated sterilization
procedure in accordance with EN ISO 17665-1 standard
requirements.
 Use type B cycles.
 Sterilization temperature: 134°C.
 Stay time: 5 minutes.
 Dry time: 10 minutes.
100 cycles
Storage times shall depend on the type of packaging used and the
storage environment. Packaging must be protected against dust,
humidity and all contamination risks.
- 19 -
2
ASPINA - DO M
3