INSTALLATION – OPERATION – MAINTENANCE
4.
FUNCTION DESCRIPTION
4.1. Function Description
After switching-on the main switch (5) into the position „I" the network indicator (3) goes on. Taking
the suction tubing (6) off the holder (2) will actuate the suction aggregate (9) and the underpressure
occurs at the suction mouthpiece (1). After repeated putting the suction tubing into the holder, the
suction aggregate turns off. When the separation vessel (11) is filled with waste products, the suction
aggregate turns off and the indicator for the separation vessel (4) filling goes on. Then it is necessary
to return the suction tubing back into the holder and to empty the separation vessel. During a longer
work, mainly with the desalivating mouthpiece, the box temperature may increase; the cooling fan is
then automatically actuated. The fan turns off automatically, when the box temperature drops
4.2. Detailed Description of the Suction Part Function
The underpressure air flows together with sucked waste products from the oral cavity through the
tubing system from the suction mouthpiece (1) at first through the inlet sieve (7), wherein the solid
impurities are entrapped. The underpressure air, together with the sucked waste products, free of
solid impurities greater than 2 mm, then flows into the separation vessel (11) wherein it is separated
from the underpressure air and entrapped into the separation vessel.
Suction air flows through the filter (18) and into the suction unit (9), from which it is pushed through
a noise muffler. Therein the output air passes through the output pre-filter (15) and the bacteriologic
output filter (14). After passing the filters, the air free of impurities is blown off into a free space under
the dental aspirator.
ASSEMBLY
5.
STORAGE AND TRANSPORT CONDITIONS
The dental aspirator is delivered ex-work packed in a transport carton, preventing the equipment
from being damaged during the transport.
For the transport, always use an original product packing, if possible.
Transport the dental aspirator in an upright position.
Protect the aspirator during transport against moisture, impurity or extreme temperatures.
The aspirator in an original packing can be stored in warm, dry and dustfree rooms.
Keep the packing material for possible future return shipments. If this is not possible,
dispose the packing material in an environmental-friendly way. The transport carton can
be disposed off as the old paper.
Mobile dental aspirator may only be transported with emptied separation vessel.
Always empty the separation vessel contents before any transport.
Storing or shipping the equipment in any conditions other than those specified
below is prohibited.
Climatic conditions for the storage and transport
Temperature –25°C up to +55°C, 24 h up to +70°C
Relative air humidity 10% up to 90 % (without condensation)
NP-DO M-37_10-2020
All manuals and user guides at all-guides.com
- 13 -
ASPINA - DO M