ASPINA - DO M
6.
INSTALLATION AND INITIAL START-UP
Before the initial start-up, remove all locking means serving for fixing the equipment
during the transport.
First actuation can be done only by professional specialist.
ANY MODIFICATION OF THIS EQUIPMENT IS FORBIDDEN!
This equipment cannot be used nearby other instruments. If this equipment is used
nearby other instruments, the equipment must be observed in order to verify normal
operations in the configuration it will be used.
Instruments may be affected electro-magnetically!
6.1. Environmental Conditions
The equipment may only be installed and operated in dry, well ventilated and dust-free rooms.
The mobile dental aspirator should be installed with regard to an easy access for the operation
and maintenance and good accessible rating plate.
The equipment should stand on a plane, sufficiently stable base (be aware of the aspirator
weight, see item 3. Technical data).
Neither the lines for the connection to the electric network, nor the suction tubings
may be cranked.
To guarantee a trouble-free operation of the aspirator, the room temperature should never drop
under +5˚C and exceed +40°C. The ideal ambient temperatures are between +10°C up to +25°C.
Most of the electric energy used by the suction aggregate (9) is converted to heat and given off
to the ambient. During a longer lasting work, above all with the desalivating mouthpiece, the box
temperature increases over 40°C causing an automatic actuation of the cooling fan. After the
room is cooled down under approximately 32°C, the fan turns off again.
6.2. Electric Connection
The dental aspirator model 230V is delivered with the safety plug. It is unavoidable to observe the
rules of the local electricity works. The network voltage and frequency must be in compliance with
the rating plate data.
In case that the equipment is fix connected to the supply of electric voltage, a disconnecting
device must be situated near the equipment.
If the equipment is connected to the supply of electric voltage by means of a plug, the socket
must be good accessible for safety reasons, so that the equipment can safely be unplugged from
the network in case of a danger.
The network distribution must be protected by maximum 10 A.
The power cord can only be replaced by maintenance personnel!
6.3. Initial Start-ip
After unpacking, put the aspirator on the floor, open the door and check whether the separator's
cover (10) with sensing electrodes is attached to the separation vessel (11). In case of need, fix the
cover to the vessel and then follow the instructions in the Chapter No. 9. Close the door and connect
the aspirator's power cord to the socket. Change the main switch (5) over into the position "I", the
indicating lamp for the network (3) goes on and the aspirator is ready for the operation.
Suction cannulas must comply with local legislative and technical requirements for
dental suction cannulas and their connection dimensions must be compatible with
suction cannulas connectors of 11mm and 16mm diameter
All manuals and user guides at all-guides.com
INSTALLATION – OPERATION – MAINTENANCE
- 14 -
NP-DO M-37_10-2020