It´s forbidden to cover all blow-holes placed on upper sides of unit.
Aspirator contains no spare power supply!
Before each switch on running unit it´s neccessary to check if door of box is closed.
8.
MAINTENANCE INTERVALS - USER / TECHNICIAN
Notice!
The operating entity is obliged to ensure that all tests of the equipment are carried out
repeatedly at least once within every 24 months (EN 62353) or in intervals as specified by the
applicable national legal regulations. A report must be prepared on the results of the tests
(e.g.: according to EN 62353, Annex G), including the measurement methods used.
Maintenance to be done
Use of antifoaming tablets
Inlet Sieve Cleaning
Lubricating seals and sliding
clamps
Disinfection of Tubings and
Separation Vessel
Cleaning and disinfection of the
exterior surfaces of the product
Replacement of output pre-filter
Replacement of output filter
Exchange of suction cannula
Replacing the filter in the
separation vessel
Perform "Repeated Test"
according to EN 62353
9.

MAINTENANCE, CLEANING AND DISINFECTION

Repair works exceeding the scope of a general maintenance may only be carried
out by a qualified specialist or by the maufacturer's customer service.
Use only spare parts and accessories permitted by the maufacturer.
Before making any maintenance, repair or cleaning, it is unavoidable to disconnect
the equipment from the network (unplug).
Allow the equipment to cool down before maintenance and service!
Reconnect the grounding lead if removed during service work to its previous
location once the work is complete.
During the preparation phase for suction cannulas, always wear EC marked
personal protective equipment (PPE).
All manuals and user guides at all-guides.com
INSTALLATION – OPERATION – MAINTENANCE
Chapter
Time interval
7.2
9.1
9.2
Every 15 days
9.3
9.4
9.5
every 3 months
9.6
max. 100 steam cleaning
9.7
9.8
Every 18 months
8
- 16 -
at need
per day
per day
at need
per annum
cycles
1x in 2 years
To be
performed
by
user
user
User
user
user
user
user
user
User
qualified
technician