INSTALLATION – OPERATION – MAINTENANCE
Cannulas are not delivered in a sterile state!
Suction cannulas must undergo complete preparation cycles before first use and
after each patient or use following instructions in Chapter 9.7
USE
7.
OPERATION
The equipment may only be operated by the trained staff!
Disconnect from the network in case of a danger (unplug) ! Hot surfaces of the
suction aggregate. Do not touch - Danger of burn.
7.1. Switching on the Dental Aspirator
Switch on the dental aspirator by turning the power-supply switch (5) in the position „I".
The aspirator is put into operation automatically by taking the suction tubing (6) with a mouthpiece
(1) off the holder (2). It works uninterruptedly until both tubings are put on the holders again or until
the separation vessel (11) is full (during an uniform operation it is filled within about 6 – 10 hours).
Suction tips holder is equipped with vacuum regulation of suction unit depends on dentist needs.
The state of filling of the separation vessel is evaluated by the separation automatics which interrupts
the suction by the aggregate (9) and filling of the separation vessel is indicated automatically with
the indicator lamp (4) going on. At this point it is necessary to put the tubings with suction
mouthpieces into the holder and switch the power-supply switch off. Afterwards open the door by
pulling the grips situated on the side walls, uncouple rubber fixing lugs from the separator cover (10)
and pull the separator's vessel out. Hang the separator's cover into the holder on the left side (8).
Empty the separation vessel content into the sewer, rinse the vessel with water, connect with the
separator's cover in a reversal order. Set the separator into a bowl (12) in the aspirator box (the bowl
is detachable). Check the connection of the cover with the vessel, seating of the separation vessel
in the bowl and close the door.
7.2. Use of antifoaming tablets
Under specific exhaust conditions, increased exhaust condensate foaming may occur due to the
switching off of the exhaust unit when the separation tank is not full. Antifoaming tables (17) must
be inserted in order to prevent this from occurring (see the basic accessories); add these tables to
the intake screen (7). The tablets will then gradually dissolve in the input screen and will significantly
decrease exhaust condensate foaming; the tablets also function as disinfection agents.
7.3. Filter use in the separation vessel
The filter in the separation vessel is an additional layer of protection for the suction unit to prevent
moisture from the foam from entering the suction unit. Foam is generated in the separation vessel
under specific suction conditions. Foam-blocking tablets are used to eliminate such foam, see
Chapter
Schedule for adding tablets: Insert 1 or 2 antifoaming tablets into the intake screen (7) at the end
of work and after cleaning and disinfecting the equipment.
Suction cannulas must undergo complete preparation cycles before first use and
after each patient or use following instructions in Chapter 9.7
Used suction cannulas intended for repeated use must be cleaned and steam
sterilized after every patient (see Chapter 9.7).
The cannulas are to be exchanged after maximum of 100 cycles of steam sterilizing.
Cannulas must be visually inspected for any signs of wear or tear before they are
used. -Never use the cannulas if it is damaged: discard it immediately.