Annual Calibration:
The AD629 had been designed to provide many years of reliable service, but annual calibration is
recommended due to possible impact on transducers.
We also required calibrating the AD629; if something drastic happens to a part of it (e.g. headset or
bone conductor is dropped on a hard surface).

How to clean Interacoustics Products

Caution:
Before cleaning always switch off and disconnect from the power supply
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
Do not allow liquid to come in contact with the metal parts inside the earphones / headphones
Do not autoclave, sterilize or immerse the instrument or accessory in any fluid
Do not use hard or pointed objects to clean any part of the instrument or accessory
Do not let parts that have been in contact with fluids dry before cleaning
Rubber ear-tips or foam ear-tips are single use components
Ensure that isopropyl alcohol does not come into contact with any screens on the instruments
Ensure that isopropyl alcohol does not come into contact with any silicone tubes or rubber
parts
Recommended cleaning and disinfection solutions:
Warm water with mild, nonabrasive cleaning solution (soap)
Normal hospital bactericides
70% isopropyl alcohol only on hard cover surfaces
Procedure
Clean the instrument by wiping outer case with a lint free cloth lightly dampened in cleaning
solution.
Clean cushions and patient hand switch and other parts with a lint free cloth lightly dampened
in cleaning solution.
Make sure not to get moisture in the speaker portion of the earphones and similar parts
AD629 Instruction for Use – English