Storing the weight value (HOLD)
– Press the HOLD key whilst the load is
on the scale.
The display flashes until a stable weight
has been measured. The display is then
frozen until the next weighing operation
takes place.
– "HOLD" and
display. The triangle with the exclama-
tion mark indicates that this frozen dis-
play is an additional facility which is not
subject to calibration.
– The HOLD function can be switched
off by pressing the HOLD key again.
0. 0
appears in the display if the scale
has no load on it, otherwise the current
weight. The "HOLD" display vanishes.
Note:
The HOLD function can also be used if the TARE function has been executed
beforehand.
Checking that the contents of the calibration counter are correct
This seca scale is calibrated. Calibration may be performed only by authorised centres.
To ensure this, the scale is fitted with a calibration counter which records any change to
data relevant to calibration. If you want to check whether the scale is properly calibrated,
proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the
scale.
– The current contents of the calibration
counter are shown on the display for a
few seconds.
– Compare the output contents of the cal-
ibration counter with the number shown
on the calibration counter sticker.
For valid calibration, both numbers must
match. If sticker and calibration counter do
not match, then re-calibration is required.
Contact your service agent or seca Cus-
tomer Service.
Model 336
appears in the digital
Both numbers must match.
15