Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and the Oxyfrog® Pulse Oximeter
Oxyfrog® Pulse Oximeter is intended for use in the electromagnetic environment
in which radiated RF disturbances are controlled.
The customer or the user of the Oxyfrog® Pulse Oximeter can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
Oxyfrog® Pulse Oximeter as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power of
transmitter (W)
0,01
0,1
1
10
100
Separation distance according
to frequency of transmitter (m)
150 KHz
to 80 MHz
3
5 .
d
=
P
V
E
1
1
0,12
0,37
1,17
3,69
11,67
56
80 MHz
to 800 MHz
3
5 .
d
P
=
E
1
0,12
0,37
1,17
3,69
11,67
800 MHz
to 2,5 GHz
7
d
=
P
E
1
0,23
0,74
2,33
7,38
23,33