Guidance and manufacture's declaration-electromagnetic immunity
Guidance and manufacturer's declaration-electromagnetic immunity
The Oxyfrog® Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The user of the Oxyfrog® Pulse Oximeter should
ensure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz) magnetic
field
IEC 61000-4-8
for all EQUIPMENT and SYSTEMS
IEC60601
Compliance
test level
level
±6KV contact
± 6 kV contact
±8KV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic
environment-guidance
Floors should be made of
wood, concrete or ceramic
tile. If the floor is covered
with synthetic material, the
relative humidity should be at
least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
53