Once finished push switch on O position and unplug.
Remove the accessories and clean.
At the end of each use, place the device in its box away from dust.
WARNING: The power supply cable plug is the element of separation from the electrical mains system: even if the units
equipped with a special on / off switch button, the power supply plug must be kept accessible once the device is in use so as to
allow a further method of disconnection from the mains supply system.
Autoclavable Silicon 8x14 mm
Vacuum (Kpa / Bar)
"PATIENT" Port
"VACUUM" Port
WARNING: NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
Footswitch control device: The equipment, on request, is provided with a footswich control device. It allows the continuous
use of the surgical aspirator. In this case, the plug of the footswitch device shall be inserted into the appropriate socket outlet
placed on the back side of the equipment.
Using the footswitch control and the flow deviator: If using equipment fitted with a flow deviator, users may direct suctioned
liquids in any of the two collection tanks provided. Flow deviator comes with two complete suction kits (2 sets of tubes, 2
antibacterial and hydrophobic filters and two conical junctions).
Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove the blu ring fitted
on the tank holder for a more comfortable insertion of the same container.
After opening the package, fully stretch the bag and then flatten it concentrically to eliminate as much air as possible.
-
Insert the bag and apply the cover to an appropriately sized reusable rigid container by pressing firmly around the entire
-
perimeter. Make sure that the system is completely sealed.
close the connector marked as "TANDEM" with the lid provided.
-
Connect the power source of the vacuum to the VACUUM port equipped with specific reusable conical fitting with "male"
-
connection.
Connect the patient tube to the PATIENT port of the cover
-
Near the footswitch control plug, there is a commutator that operates the device
directly using the mains or by using the footswitch control.
When the switch is turned on (green light) and the commutator is in position I, the
device works directly using the mains.
When the switch is turned on (green light) and the commutator is in position II, the
device works via remote control (footswitch control)
When the switch is turned on (green light) and the commutator is in position 0, the
device is powered but not in operation
20
Antibacterial Filter
Vacuum Regulator Knob
ON / OFF Switch
Conical Fittings
Collection Jar in Autoclave PC
GB