Gima SUPER VEGA Mode D'emploi page 16

Aspirateur - 2 x 2 l - sur chariot avec interrupteur à pédale
Masquer les pouces Voir aussi pour SUPER VEGA:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

GB
ANTIBACTERIAL AND HYDROFOBIC FILTER
(for versions equipped with footswitch control)
The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit. It should be changed
immediately if it becomes wet or if there is any sign of contamination or discolouration. If should also be changed if the unit is used
with a patient whose risk of contamination is unknown. Don't use the suction unit without the protection filter. If the suction
unit is used in an emergency or in a patient where the risk of contamination is not know the filter must be changed after each use.
On request, are available versions with disposable collection system (FLOVAC® 2000ml) composed by a reusable polycarbonate
vessel and polyethylene disposable bag (collection liner for fluids suction canisters).
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-
infections. Don't use after lapse of the sell-by date
packaging. GIMA declines any liability for injury to the patient correlated to the deterioration of the above-
Check the expiry date on the original packaging of the suction catheter and check the integrity of the sterile
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO
10993-1 standards on material biocompatibility.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond this
limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you to
change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube.
Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must
be replaced if there are visible signs of decay of the material constituting the said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the
said component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse.
The component must be replaced if there are visible signs of decay in the material constituting the said component.
Service life of the device: More than 850 hours of operation (or 3 years) in accordance with the standard conditions of testing and
operation. Shelf life: maximum 5 years from the date of manufacture.
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating
substances;
Disconnect the tank from the device and remove the said container from the support of the device.
Separate all the parts of the cover (overflow device, washer).
Disconnect all tubes from the jar and the protection filter
Wash each part of the container from secretions under cold running water and then clean every single part in hot
water (temperature not exceeding 60°C)
Once again, carefully wash each single part using, if necessary, a non-abrasive brush to remove any deposits. Rinse
with hot running water and dry all parts with a soft cloth (non-abrasive). It is possible to wash with commercial
disinfectants by carefully following the instructions and dilution values supplied by the manufacturer. After cleaning,
leave the parts to dry in an open, clean environment.
Dispose of the aspiration catheter according to that provided by local laws and regulations.
The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed
60°C). After cleaning, leave the parts to dry in an open, clean environment.
When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
ACCESSORIES SUPPLIES
DESCRIPTION
N°2 COMPLETE ASPIRATION JAR 2000cc
CONICAL FITTING
TUBES SET 8 mm x 14 mm
FOOTSWITCH CONTROL
EUROPEAN POWER SUPPLY CORD
mentioned sterile packaging due to handling of the original packaging by third parties.
CLEANING OF ACCESSORIES
RE 210351/01
RE 210355/03
17
CODE
RE 210420
SP 0121
Cod. 52130
SP 0020

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

28192

Table des Matières