Weight value incorrect
Cannot use Full Capacity
Not linear
Off center loading error
Battery will not charge
11. WARRANTY
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the delivery date.
During this period, GIROPÈS SL will take charge of repairing the scale.
This warranty does not cover defects or damaged caused by misuse, overloading or improper installation contrary to the
recommendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.
30
Calibration error, Recalibrate.
Linearity error, set Linearity.
Unit
calibrated
with
weight.
Balance not level.
Obstruction between sample and co-
ver.
Wrong unit of weight displayed.
Over load Stops hitting pan support or
hitting bottom of load cell.
Shipping screw not removed if appli-
cable.
Electronic problem on A/D.
Parameters set incorrectly.
Load cell or mechanics damaged.
Overload stops hitting too soon.
Load cell or mechanis damaged.
A/D damaged.
Adjust mechanics.
Overload stops not correct.
Load cell damaged.
∏
Mains voltage not present or too low.
Charging circuit failure.
Battery failure.
Calibrate again, paying special atten-
tion to the mass used, the stability of
inaccurate
the scale, and the weighing units re-
quired.
If linearity can be set by the user it will
be described in the user manual.
Check pan is installed correctly.
Verify installation is acceptable.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-
tion.
Check the weighing units used.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-
tion.
If linearity can be set by the user it will
be described in the user manual.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-
tion.
Verify the batteries are rechargeable
types.
Check power supply voltage is correct.
BAR model