Connecting the pressure gauge
• Open the cap on the PEP S connecting element.
 CAUTION! Risk of infec-
tion due to contamination! If
the connection tubing is con-
nected directly to the
PARI PEP S System, germs
may build up in the tubing.
Since the connection tubing
cannot be disinfected, use the
pressure gauge exclusively
with the extra pressure gauge
connector line and the tubing
adapter, in order to avoid a risk of infection.
Push the pressure gauge connector line onto the pressure
gauge attachment point on the PEP S connecting element.
• Insert the corresponding end of the connection tubing in the
pressure gauge.
• Connect the connection tubing and the pressure gauge con-
nector line using the tubing adapter.
3.2 Performing treatment
All the safety instructions and warnings in these instruc-
tions for use must have been read and understood before
any treatment is carried out.
Proceed as follows in order to carry out the treatment:
• Before starting the treatment, ensure that all parts are firmly
connected to each other.
• Make sure that the expiratory resistance is adjusted cor-
rectly [see: Adjusting expiratory resistance, page 14].
If the breath resistance seems too high or too low to you
during the therapy session, stop the session and consult
your doctor or therapist.
– 16 –