Cleaning and disinfecting
Manual cleaning
EQUIPMENT
The method has been validated in Europe using:
– pH-neutral, enzymatic cleaner:
Bode Bomix
– Drinking water temperature of about 15 °C
– Application time: 10 minutes
PROCEDURE:
CAUTION
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Observe the mixing proportions indicated and the spe-
cified treatment time.
• Ensure that all components are completely submerged in
the solution for the whole of the treatment time. There
must not be any air pockets or bubbles.
• Clean all individual parts with a solution prepared according
to the manufacturer's instructions.
In case of visible soiling, use a medium-soft brush (e.g. a
toothbrush) which is used exclusively for this purpose.
If the recommended application period is exceeded signi-
ficantly, the plastic parts may take on the smell of the me-
dium used.
RINSING:
• Rinse all individual parts thoroughly under running drinking
water at approx. 15 °C for 3 minutes.
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent surface, and
allow them to dry completely.
– 27 –
®
plus (concentration: 0.1%)