Technical Specifications; General; Characteristics; Dc Voltages - Metrix MX 26 Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 4
Chapter II
6.

TECHNICAL SPECIFICATIONS

6.1.

General

Only the values assigned tolerances or given limits are guaranteed values. Values without
tolerances are given for information only, without guarantee (standard NFC 42670). The
measurement errors must be considered in an environment of reference temperature (refer to §
6.2.11).
It is essential that all measuring instruments are regularly calibrated.
6.2.

Characteristics

Accuracy is ± [% reading (R) + number of digits (D)].
{Accuracy : ''n %R + n D'' means ''n % of the reading + n digits'' }

6.2.1. DC voltages

Ranges
Selector switch
position
ADP
500 mV
DC
DC
V
5 V
DC
DC
50 V
DC
500 V
DC
1000 V
DC
(∗) Acceptance permanent max. voltage
Number of points :
Range selection :
Common mode rejection :
Serial mode rejection :
Intermittent signal sounds and « OL » on the display in case of range overpassing.
For measurements performed on alternative signals, the selected range must tally with the max.
value of the signal peak.
Additional error in Pk +/- mode for a pulse width of ≥ 1 ms : 1 % R ± 50 D

6.2.2. AC voltages (AC and AC+DC)

Switch
Input
Range
position
impedance
5 V
AC
50 V
AC
VLOW Z
500 kΩ
500 V
AC
750 V
AC
10 MΩ
ADPAC+DC
500 mV
AC+DC
//100 pF
5 V
AC+DC
11 MΩ
50 V
AC+DC
VAC+DC
500 V
AC+DC
10 MΩ
750 V
AC+DC
(∗) Acceptance permanent max. voltage
Specifications applicable from 10 % to 100 % of the range for a sinusoidal signal in band 40 Hz
to 20 kHz and from 20 % to 100 % of the range in band 20 kHz to 100 kHz.
Number of points :
Range selection :
Common mode rejection :
Additional error according to crest factor :
Intermittent signal sounds and « OL » on the display in case of range overpassing.
5000-count portable digital multimeter, with current inputs
Input impedance
Accuracy
0,3 % R + 2 D
10 MΩ
11 MΩ
10 MΩ
5 000
automatic or manual for the 5 V, 50 V,
500 V, 1000 V ranges
at 50 and 60 Hz, better than 120 dB
at 50 and 60 Hz, better than 60 dB
Accuracy
Protection
(*)
40Hz to 1kHz
1 to 4 kHz
4 to 20 kHz 20 to 50 kHz 50 to 100 kHz
1 % L + 3 D
2 % L + 3 D 3 % L + 3 D
± 1100 Vpk
1 % L + 3 D
775 Vrms
-
± 1100 Vpk
1.5 % L + 3 D
-
775 Vrms
1 % L + 3 D
2 % L + 3 D 3 % L + 3 D
± 1100 Vpk
1 % L + 3 D
775 Vrms
-
5 000
automatic or manual for the 5 V, 50 V,
500 V, 750 V ranges
at 50 and 60 Hz, better than 60 dB
0.5 % for a crest factor of 2 to 3
1 % for a crest factor of 3 to 6
(∗)
Resolution
Protection
± 1100 V
0.1 mV
PEAK
DC
775 Vrms
1 mV
DC
10 mV
DC
100 mV
DC
1 V
DC
Resolution
1 mV
5 % L + 3 D
10 mV
100 mV
-
-
-
1 mV
-
-
-
0.1 mV
1 mV
5 % L + 3 D
10 mV
100 mV
-
-
-
1 V
AC+DC
AC
AC
AC
AC
AC+DC
AC+DC
AC+DC
AC+DC