Chapter II
3.5.2. Battery self-test
When the BAT indication appears on the display, the instrument still has approximately 12
hours of operation, but specifications can no longer be guaranteed.
Replace the battery.
3.5.3. Replacing the battery or fuses
Caution !
Disconnect test leads from any circuit under test, turn the meter off
and remove test leads from the input terminals .
Use the following procedure:
1 - Disconnect test leads from any input terminals.
2 - Using an appropriate tool, slide off the case bottom of the instrument.
3 - Replace the battery or fuse (respect the value and the type).
4 - Replace the removable part.
F1 0.63 A HBC 18 kA / 600 V
Fuses must be replaced
by fuses of the same
rating and type.
Fitting the fuses :
Those fuses are correctly fitted ; those ones are not.
3.5.4. Cleaning
Clean the instrument using a damp cloth and soap. Never use abrasive or solvents.
3.5.5. Storage
To guarantee the measurement accuracy, after a long period storage in extreme
environment conditions, wait enough time for the instrument to acclimatise to the working
environment conditions (see environment specifications).
5000-count portable digital multimeter, with current inputs
F2 10 A HBC 50 kA / 600 V
6LF22
9 V BATTERY
21