Technical Specifications; General; Characteristics; Dc Voltage (Autorange) - Metrix MX 350 Notice De Fonctionnement

Masquer les pouces Voir aussi pour MX 350:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 3
5.

TECHNICAL SPECIFICATIONS

5.1. General

Only the values assigned tolerances or the limits
declared constitute guaranteed values. The values
without any tolerance are given as indications.

5.2. Characteristics

The accuracy is ± [% of the reading (R) + number of
representation units (digits or D)] in the reference
conditions (see appendix).

5.2.1. DC voltage (Autorange)

Range
Measurement range
Resolution
400 V
0.2 V to 399.9 V
0.1 V
600 V
400 V to 600 V
1 V
Protection against overloads: 660 Vrms

5.2.2. AC voltage (Autorange)

Range
Meas. range
Frequency
400 V
0.5 V to 399.9V 50 .. 500 Hz
600 V
400 V to 600 V
50 .. 500 Hz
MX 350: Input impedance: 1 MΩ
MX 355: Input impedance: 10 MΩ
Protection against overloads: 660 Vrms

5.2.3. DC current (Autorange)

(MX 355)
Range
Measurement range
Resolution
40 A
0.10 A to 39.99 A
0.01 A
400 A
40.0 A to 400.0 A
0.1 A
Protection against overloads: 600 Arms

5.2.4. AC current (Autorange)

(MX 350)
Range
Meas. range
Frequency
40 A
0.05A to 39.99A 50 .. 60 Hz
60 ... 500 Hz
400 A
40.0A to 400.0A 50 .. 60 Hz
60 ... 500 Hz
Protection against overloads: 600 Arms
(MX 355)
Range
Meas. range
Frequency
40 A
0.05A to 39.99A 50 .. 500 Hz
400 A
40.0A to 400.0A 50 .. 500 Hz
Protection against overloads: 600 Arms
28
Chapter II
Accuracy
1% R +2 D
1% R +2 D
Resol.
Accuracy
0.1 V
1.5%R +5 D
1 V
1.5%R +5 D
Accuracy
2.5% R +10 D
2.5% R +10 D
Resol.
Accuracy
0.01 A
1.9%R + 5 D
2.5%R + 5 D
0.1 A
1.9%R + 5 D
2.5%R + 5 D
Resol.
Accuracy
0.01 A
2%R +10 D
0.1 A
2%R +10 D
MX 350 / MX 355

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mx 355

Table des Matières