E. DISINFECTING YOUR PARI
REUSABLE NEBULIZER
1. Clean the nebulizer as described in Section
D. For low-level disinfection every other
treatment day, soak all parts of the PARI
Reusable Nebulizer (except tubing and mask)
in a solution of 1 part distilled white vinegar
and 3 parts hot tap water for 1 hour. You can
substitute respiratory equipment disinfectant
(like Control III
) for distilled white vinegar
®
(follow manufacturer's instructions for
mixing).
2. Rinse all parts of the PARI Reusable
Nebulizer thoroughly with warm tap water
and dry with a clean, lint-free cloth. Discard
the vinegar solution when disinfection is
complete.
3. The PARI Reusable Nebulizer parts (except
tubing and mask) may also be disinfected by
boiling them in clean pot of water for a full
5 minutes.
4. Hospitals and clinics can sterilize the PARI
Reusable Nebulizer parts (except mask and
tubing) by autoclaving (a maximum of 277° F
or 136° C), follow autoclave manufacturer's
instructions.
5. Dry parts on a clean, lint-free cloth.
CONTROL III
®
IS A REGISTERED TRADEMARK OF MARIL
PRODUCTS, INC.
NOTE: FOR PATIENTS WITH CYSTIC FIBROSIS, THE CF
FOUNDATION RECOMMENDS BOILING FOR A FULL 5
MINUTES TO DISINFECT.
PARI Respiratory Equipment, Inc.
2412 PARI Way, Midlothian, VA 23112 USA
Tel: 800.FAST.NEB (327.8632) and 804.253.7274
|
Fax: 800.727.4112
Product.Info@PARI.com
Online at: www.PARI.com
6 Questions? 1.800.FAST.NEB (327.8632) | PARI.com
F. WARRANTY AND RETURN
INFORMATION
PARI reusable nebulizers include a 6 month
limited warranty for home care use.
If you have any problems and need to return
your PARI reusable nebulizer, contact PARI
Patient Connection at 1.800.FAST.NEB
(327.8632)
Please be sure to clean and disinfect your PARI
reusable nebulizer and place it in a plastic bag
before returning.
© 2015 PARI Respiratory Equipment, Inc.
023D0222 Rev A 10/15