International regulations (cont.)
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
54
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
a
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the breast pump is used exceeds the applicable RF compliance level above, the breast pump should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the breast pump.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
b
Table of frequencies of portable and mobile transmitters for which the recommended
separation distance is 30 cm (12 inches):
Band (MHz)
380 - 390
430 - 470
704 - 787
800 - 960
1 700 - 1 990
2 400 - 2 570
5 100 - 5 800
WARNING: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the breast pump including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
Service
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 17
GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5
GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS
Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7
WLAN 802.11 a/n