9 Faults

Warning:
-
When carrying out work on the device, please disconnect the power supply.
-
Repairs must be carried out by a skilled electrical fitter.
Schedule 1, faults
Probleem
Ion balance cannot
be adjusted to 0
Ion balance does
not change when I
turn the control
knob.
Ion balance
changes but the
control range is
small or limited to 1
polarity.

10 Repairs

Warning:
-
When carrying out work on the equipment, disconnect the power supply.
-
Repairs must be carried out by a skilled electrical fitter.
To order spare parts please check the spare parts list.
In order to return the product, please apply for an RMA form by e-mail at service@simco-ion.nl.
Pack the device properly and clearly state the reason for return.
ExternalBalanceControlKit_MPM_UM_9752080710_NL_D_GB_F_V2_0
Mogelijke oorzaak
Neutralising equipment is fouled
DIP switch 2 in the MPM is not in
the "ON" position
A capacitively coupled rod is
used as neutralising equipment
Wires incorrectly connected
Neutralising equipment is fouled
Points of the neutralising
equipment are worn
Oplossing
Clean the neutralising equipment
Turn DIP switch 2 into the "ON"
position
Use a directly coupled or resistor
coupled rod as neutralising
equipment
Check the wiring
Clean the neutralising equipment
Replace the neutralising
equipment
38