6.1.3 Battery check

The battery indicator on the display shows the state of battery.

6.1.4 Switch to Ion Balance measurement

The FMX-003 can easily be used to measure ion balance voltage (residual voltage)..

6.1.5 Placing the ion balance plate

1 Remove the plate assembly (depress the IB plate release buttons on either side of the
fieldmeter).
2 Insert the plate assembly into the top end of the fieldmeter (the end with the Simco-Ion logo)
until it clicks into place.
3 Connect the instrument with earth, using the supplied grounding lead (when not already in
place).
- Press the [IB] button, to turn on the ion balance measuring circuit.
The text [IB] is visible on the display.

6.1.6 Zeroing

During IB measurement the indication should be zero when connecting the IB balance plate to
ground.

6.1.7 Measurement

Warning:
For reliable measurements the relative humidity must not exceed 60%.
Note
-
To obtain proper measurements, earthing of the measuring instrument is essential.
-
The buzzer sounds continuously, when a value higher than 220 volts is displayed
(overrange). Stop measurement in this case.
Place the meter at a defined distance (for example 50mm) from the top of the neutralizing
equipment to be measured.
The measuring instrument indicates the voltage measured (V) and the polarity.
ExternalBalanceControlKit_MPM_UM_9752080710_NL_D_GB_F_V2_0
Replace the battery