6. Measuring and adjusting ion balance

6.1 Measuring ion balance with FMX-003

− Read manuals from MPM and neutralising equipment.
− Make sure that the equipment is properly earthed.
Earthing is needed to ensure proper operation and to avoid electrical shocks upon contact.
− Electrical installation, maintenance and repairs shall be made by a skilled electrical engineer.
− When carrying out work on the equipment: de-energise the equipment.
Attention:
− Before adjusting the Ion Balance be sure the neutralising equipment is in clean
condition, see manual.
Clean emitters as instructed below.

6.1.1 Earthing the measuring instrument FMX-003

The case of the FMX-003 is made of a conductive resin which provides an earth reference for
the measuring circuit. To ensure accurate measurement the instrument must be earthed using
the supplied grounding lead.

6.1.2 Switch-on with continuous mode

- Press and hold the red button [POWER] for at least three seconds when turning the FMX-003
on (Three led bars, the A-OFF sign and three beeps indicates that the instrument is switched
in continuous mode)
The FMX-003 will NOT turn itself off automatically after five minutes. When the measuring
instrument is switched on, the graphic and numerical indications and the battery symbol are
displayed.
When the instrument is switched in continuous mode, the red indicating LED's will be turned off
automatically after one minute to reduce the consumption of the battery.
ExternalBalanceControlKit_MPM_UM_9752080710_NL_D_GB_F_V2_0