These assembly instructions describe:
•
fitting ferrite beads to the connecting cables for the power
supply and to the battery of the Accuvac Rescue suction pump
•
fitting a capacitor to the charging interface of the
Accuvac Rescue suction pump
The ferrite beads and the capacitor are components of the
Accuvac Rescue suction pump. Always also read the
Accuvac Rescue service and repair instructions.
Installing the ferrite bead for the power supply
Ferrite bead
for power
supply with
cable guard
1. Place the ferrite bead for the power supply around the
connecting cables for X101 and X100 and squeeze it together
until it clicks.
2. Make sure the adhesive surface on the housing is free of dust,
oils and grease.
3. Remove the adhesive film from the cable guard and glue the
cable guard to the housing underneath the ferrite bead.
4. Remove the adhesive film from the cable holders.
5. Position the cable holders at the ends of the ferrite bead and
glue them on.
6. Insert the connecting cables through the cable holders.
X101
X100
Ferrite bead
for battery
Accuvac Rescue
EN
5