11
ENGLISH
OPERATION
Functions description
Mute on/off
There is a acoustic signal when SpO2 and Pulse Rate are detected. You may press the MUTE button
to keep the silence, and the MUTE light
will be on.
Volume
Press VOLUME
button to choose the different sound level
Feed
Press FEED
button only when the battery power is in good condition to skip one line of the paper
Print
press PRINT
button only when the battery power is in good condition to print out the last 6 data stored
in the memory
Battery Box
locates at the back of the appliance which can store 4 pieces AA batteries
AC Adapter Connector
locates at the left bottom side of the appliance where the AC adapter can be connected
Sensor Connector
locates at the right bottom side of the appliance where the connector of the sensor should be plugged
in. To remove the sensor, pull out the sensor from the appliance.
- New alarm setting will be saved automatically. When running out of batteries or being disconnected
with power, the alarm setting will return to default alarm settings.
- When the detected values are beyond or below the set figures, the alarm will be on.
- Print & Feed button can be used only on the condition of full scale or pulse search bar.
Signal Display
░ ░ ░ ░ : Pulse Search Bar, indicating the strength of the pulse and patient signal. If it is in correct
detecting condition, the search bar should have 4 scales running bar at least.
Prb: Sensor disconnected with the appliance
Pot: Patient disconnected with the appliance
Lo: Weak signal from patient
SpO2 Max/Min: alarm values of SpO2
Pulse Max/Min: alarm values of Pulse.
Indicative Lights:
- TIME light on, it indicates the time setting is in function
- MEMORY light on, it indicates the memory function is on
- MUTE light on, it indicates the oximeter is working in silence.
- POWER & LOW BATTERY light, if the light in RED, it indicates the power is on. When the light turn
GREEN, it indicates the power of battery is low. When the GREEN light on, it could still work for half to
one hour. Also it is the time to consider replacing new batteries.