The Parker HPN2 range of nitrogen gas generators employ robust, field proven technology to produce high purity nitrogen for life science,
chemical analysis and spectroscopy applications. Flow rates range from 5 L/min (0.18 cfm) to 12.5 L/min (0.44 cfm), with purities from <10ppm
to <0.5% at an outlet pressure of 5 barg (72.5 psig / 0.5 MPa / 500 kPa).
The generators provide a continuous stream of high purity nitrogen from a single 'plug & play' unit. Models are available with and without an
integral oil free compressor

2.1 Technical Specification

This specification is valid when the equipment is located, installed, operated, and maintained as specified within this user guide.
Outlet Parameters
Flow
(1)
Set Pressure
(2)
Purity
content
(O
)
2
(3)
Dewpoint
(@ 5bar)
Particulates
Inlet Parameters
Integral Compressor
Air Quality
Air Pressure
Minimum Inlet Air Temperature
Maximum Inlet Air Temperature
Air : N
Ratio (@ 9 barg)
2
Peak Flow (@ 9 barg)
Port Connections
(4)
Air Inlet
Gas Outlet
Drain
Electrical Data
Supply Voltage
Connection type
Current
(5)
Power
(6)
Fuse
Supply Voltage
Connection type
Current
(6)
Power
(7)
Fuse
(1) If you are using the generator's internal Mass Flow Controller, please note that the pressure at the N2 MFC Outlet port will be approximately 0.5 bar (7.3 psi) below the set
pressure.
(2) The specified purities are referenced to 20
monitored in critical applications.
(3) The Pressure Dew Point (PDP) is the temperature at which the water vapour within the pressurised gas condenses in to a liquid.
(4) Using the 1/4 BSPP - 1/4 tube fittings provided.
(5) The power consumption in standby mode and economy mode is 3W for all models.
(6) Fuses are Anti surge (T), 250V, 5x20mm HBC, Breaking capacity 1500A @250V, IEC60127, UL R/C.
HPN2
Units
5000
L/min (cfm)
5.0 (0.18)
bar g
(psig / MPa / kPa)
--
<10ppm
o
o
PDP
C (
F)
ISO 8573-1: 2010
-
No
3.2.1
ISO 8573-1: 2010
bar g
(psi g / MPa / kPa)
o
o
C (
F)
o
o
C (
F)
--
10.9
L/min (cfm)
110 (3.89)
-
1/4"
-
-
Vac
-
A
0.54
W
88
A
5
Vac
-
A
1.15
W
89
A
5
o
C ambient temperature. A reduction in purity may be experienced with increased temperature, it is recommended that the purity is
HPN2
HPN2
HPN2
5000C
7000
7000C
7.0 (0.25)
(72.5 / 0.5 / 500)
<1000ppm
Yes
No
Yes
N/A
3.2.1
N/A
8 - 9.9
(116 - 143.6 / 0.8 - 0.99 / 800 - 990)
--
5.0
--
--
110 (3.89)
--
N/A
1/4"
N/A
207 - 253 V 50/60Hz
IEC 60320 - C14
3.15
0.54
3.15
698
88
698
5
5
5
104 - 127 V 60Hz
IEC 60320 - C14
5.2
1.15
5.2
596
89
596
8
5
8
HPN2
HPN2
10000
10000C
10 (0.35)
5
<2000ppm
-40 (-40)
Class 2
No
Yes
3.2.1
N/A
15 (59)
25 (77)
3.8
--
110 (3.89)
--
1/4"
N/A
1/8"
1/4"
0.54
3.15
88
698
5
5
1.15
5.2
89
596
5
8
HPN2
HPN2
12500
12500C
12.5 (0.44)
<0.5%
No
Yes
3.2.1
N/A
9 - 9.9
(130.5 - 143.6 / 0.9 - 0.99
/ 900 - 990)
3.63
--
129 (4.56)
--
1/4"
N/A
0.54
3.15
88
698
5
5
1.15
5.2
89
596
5
8
3