Storage Conditions

FastPack
IP System Analyzer
®
Storage Temperature:
Humidity:
Stacking Limits:

Statement of Compliance

FCC, Sec 15
EN 61326-1
IEC 1000-3-2
IEC 1000-3-3
IEC 1000-4-2
IEC 1000-4-3
IEC 1000-4-4*
IEC 1000-4-5
IEC 1000-4-6*
IEC 1000-4-11
Low Voltage Directive
EN 61010-1
EN 61010-2
UL 61010-1
CSA 22.2 NO. 61010-1
CSA 22.2 NO. 61010-2
*IEC 100-4-4 and IEC 1000-4-6 were passed under criterion C.

Disposal

This product contains recyclable materials. Do not dispose of this product as unsorted
waste. Please contact your local dealer or Qualigen, Inc. for disposal instructions.

Principle of Operation

Sample is added to the FastPack
reagents, in sealed chambers, necessary to perform the desired test. The pack label contains a bar code
with all necessary information required by the analyzer to run the test.
The FastPack
IP System analyzer performs the test by automatically mixing and moving the sample and
®
reagents within the pack. The sample and reagents are moved from one chamber to another by applying
uniform pressure to the compartment by means of internal pressure pads extended from the analyzer. The
pressure pads are driven by compressed air supplied by a small air compressor in the analyzer.
The FastPack
IP System analyzer is capable of running both sandwich, and competitive formatted
®
immunoassays. For the sandwich type assays, the chemical principle is as follows: A sample of unknown
analyte concentration is mixed with an excess amount of known concentration of capture and labeled
antibody solution. This mixture is incubated for a preset time to allow the capture antibody and the labeled
antibody to bind to the analyte in a sandwich format. This mixture is then brought into contact with the
coated paramagnetic particles, which bind to the capture antibody (and thus the analyte). The amount of
labeled antibody bound to the paramagnetic particles is directly proportional to the analyte concentration
in the sample.
68                                                                                                                                                                                                                                                                                                                                                                                                                             6 5000159   R ev.   0 07   ( 06/15)  
15 °C (59 °F) to 32 °C (90 °F)
10% to 80% relative humidity
Not to exceed four high
Class A Radiated Conducted Emissions
Class A Radiated Conducted Emissions
Powerline Harmonics Test
Powerline Flicker Test
ESD Immunity
Radiated Susceptibility
EFT Immunity
Lightning Surge Immunity
RF Common Mode Immunity
Voltage Dips & Short Interruptions
72/23/EEC
Safety Requirements
via the injection port. The FastPack
®