Setting Dilution Mode
Dilution Mode
This mode enables the analyzer to calculate and output results for hCG samples that have been diluted
prior to testing. This affects hCG tests only. The analyzer outputs the calculated result based on the
operator-selected dilution factor. A "D" or "X:" will be printed to the left of the result if a sample was tested
in Dilution Mode. "D" indicates that a sample was tested at a 1/100 dilution; "X" indicates that a sample
was tested at a 1/10,000 dilution. "D" or "X" will likewise appear on the second character row of the
analyzer screen under the same testing conditions. With Dilution Mode disabled, the analyzer will
immediately initiate an hCG test with no dilution prompts, and results will be output with no dilution
correction, and no indication that the tested sample had been diluted.
Setting Dilution Mode
Enable
D ilution
M ode
-‐
h CG
Action
Press
R ED
b utton
s everal
times
t o
s croll
t hrough
t he
menu
t ill
y ou
s ee
" Set
Dilution
O N"
65000159
R ev.
0 07
( 06/15)
3 1
Instrument
This
i s
n ow
y our
1 :100
diluted
p atient
s ample
t hat
you
w ill
r un
t hrough
t he
FastPack®
I P
S ystem
analyzer.
To
c reate
a
1 :10,000
diluted
s ample,
r epeat
t he
above
s teps
u sing
y our
1:100
d iluted
s ample
a s
your
p atient
s ample.
Instrument
"Set Dilution ON"
Information
Information
Diluting a Sample
Illustration
Illustration