Other Information
Upgrading Flash Memory
The FastPack
IP System analyzer is equipped with a Smart Media Card Reader that is used to install
®
periodic software upgrades to the system. As software upgrades become available, Qualigen, Inc. will
provide a Smart Media Card containing the upgrade ready to be installed on to your FastPack IP System.
Downloading Data
The FastPack
IP System (when COM1 is configured for a computer) allows for the output of test results to
®
a personal computer (PC). Data is automatically output, through an RS232 connection found on the back
of the FastPack
IP System analyzer, at the completion of each test.
®
To obtain data, connect a serial cable from the FastPack
and configure COM1 on the analyzer for a computer.
The output data is ASCII text with the following comma delimited fields:
Result Value (numerical or pass/fail for calibrator or controls)
Units
Date
Time
Instrument Serial Number
SW Revisions
Test Type
The communications protocol will be:
Serial Comm RS232
Baud Rate 9600
Parity None
Data bits 8
Stop bits 1
Recall last result sends this information out to PC for the last test run.
FastPack
IP System Product Specifications
®
FastPack
®
IP System Analyzer
Ambient Operating Temperature:
Operating Humidity:
Operating Altitude:
Pollution Degree:
Installation Category:
Display Output:
Data Output
Field Service Options:
Electrical Requirements:
Power Consumption
Size:
Weight:
This product is intended for indoor use only.
65000159
R ev.
0 07
( 06/15)
6 7
IP System analyzer RS232 port to the computer
®
15 °C (59 °F) to 32 °C (90 °F)
10% to 80% relative humidity
≤ 2000 m (6562 ft.)
2
II
2 line alphanumerical display
RS232, 9600 baud, ASCII text
ROM card firmware upgrade
100-240 ± 10% VAC, 1.9A
100 Watts (maximum)
13 x 9 x 12 inches (33 x 23 x 30 cm)
28.64 lbs (12.99 kg)