Troubleshooting the FastPack
Important Note
Proper performance of the pipette is critical to obtain accurate test results. If at any time, you note the
pipette plunger sticking, or if plunger operation feels rough, replace with a new pipette.
Problem  
Pipette   d oes   n ot   d raw   u p   o r   d ispense  
any   s ample  
Sample   i s   l eaking   f rom   t he   i njection  
port.  
 
65000159   R ev.   0 07   ( 06/15)                                                                                                                                                                                                                                                                                                                                                                                                                               6 1  
IP Injection Port and Pipette
®
Cause  
Pipette   t ip   m ay  
not   b e   p roperly  
seated.   T he   w hite  
piston   i nside   t he  
pipette   t ip   s hould  
move   u p   a nd  
down   w hen   t he  
plunger   i s  
depressed.   I f   i t  
does   n ot,   t he   t ip   i s  
not   p roperly  
seated   o r   t he  
piston   i s   w arped.  
The   p ipette   i s   n ot   p roperly   s eated   i n   t he  
injection   p ort.  
 
Action  
Eject   t he   p ipette   t ip.   S eat   a   n ew   p ipette  
tip   p roperly.   R efer   t o   t he   F astSteps  
(proper   t echniques   f or   s ampling).  
 
Discard   t he   l eaking   F astPack
with   a   n ew   u nused   F astPack
inserting   t he   p ipette   t ip   i nto   t he  
injection   p ort,   e nsure   t hat   t he   t ip   i s   f ully  
inserted   a ll   t he   w ay   i nto   i njection   p ort  
so   t hat   i t   i s   p roperly   s eated.   A fter  
injecting   t he   f luid   a nd   e jecting   t he  
pipette   t ip,   m ake   s ure   y ou   d o   n ot  
inadvertently   p ull   o ut   t he   t ip.   T his   w ill  
cause   t he   p ipette   t ip   a nd/or   t he   w hite  
internal   p iston   t o   b e   u nseated   a nd  
potentially   a llow   l eakage   o f   s ample.  
Contact   S ystem   S upport   8 77-­‐770-­‐6127.  
 
®