• Do not wash, soak, or wipe off the handpiece with/in the oxidation potential solution (strong acid, super acid solution) sterilized solution, strong
alkaline solution or functional water.
• The handpiece is kept in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. Moreover, air does not include a dust,
salt and sulfur.
• Do not touch the handpiece just after the autoclave. It may very hot.
• NSK do not recommend you to sterilize on plasma sterilization or EOG sterilization.
6. Changing Turbine Cartridge
1) Mount a dummy bur in the chuck.
2) Turn the Wrench counter clockwise and remove the Head Cap by Head Cap Wrench.
3) Push up the dummy bur to remove Cartridge.
4) Clean the head interior. (Wipe extra oil after lubrication)
5) Align the Alignment Pin and insert Cartridge. (Fig. 8)
6) Turn the Wrench clockwise and mount the Head Cap. (Fig. 9)
* Replacement Cartridge: SX-MU03 Mini Cartridge Order Code P1001250
SX-SU03 Standard Cartridge Order Code P1002250
Arraignment Slit
Arraignment Pin
Cartridge
Head Cap
Head Cap
Wrench
Fig. 9
13