atmospheric pressure:
Inside the supplied carrying bag:
temperature:
maximum relative humidity:
atmospheric pressure:
N.B. Disconnect the cables, before storing the device into its carrying bag. If not,
the cables could bend excessively near the connectors. It could severely damage
the cables.
Troubleshooting
Any type of work on MIO-CARE must be carried out exclusively by the
manufacturer or by an authorized dealer. In any event, any presumed
malfunction of MIO-CARE must be verified before sending the device to the
manufacturer.
Here below are some typical situations:
−
MIO-CARE cannot be turned on:
•
recharge the batteries and try to switch it on again; if it still does
not switch on, check that the power supply socket is working.
•
Check that the ON/OFF button has been pressed correctly (held
down for at least one second).
MIO-CARE does not transmit electric impulses or transmit low
−
intensity:
•
check that the cable jacks have been inserted in the electrodes and
that the plastic protection has been removed from the electrode.
•
Check that the cables have been connected correctly (connector
well inserted in the device).
•
Check that the electrodes have been connected correctly, are in
good conditions and correctly positioned according to user manual
instructions.
MIO-CARE switches off during operation:
−
•
recharge the batteries and start the treatment again. If the
problem persists, contact the manufacturer.
Stimulation is not comfortable:
−
•
check that intensity is not too high and decrease intensity if
needed.
•
Check the electrodes position: follow user manual and electrodes
Position manual instructions to avoid that electrodes are too close.
•
Check
manufacturer): electrodes too small can cause uncomfortable
stimulations.
IACER Srl
the
sized
electrodes
110
from 700 to 1060hPa
from -10°C to +55°C
from 10% to 90%
from 700 to 1060hPa
section
(suggested
by
the
MNPG95-05