Electrodes' positioning and intensity levels
The electrodes have to be positioned to form a square over the painful zone by
using the channel 1 and 2 as shown above in Figure 1 (red or black up or down
are not important for the therapy purposes, follow the indications in the
Positions manual). The intensity should be adjusted to a level between the
thresholds of perception and pain: the maximum intensity level is the moment in
which the muscles surrounding the treated area begin to contract; over this limit
the stimulation does not become more effective, just more irritating, so it is best
to stop before that point.
Programs specifications
Medical
Prg
prg.
Yes/No
1
Yes
Fast TENS
Endorphinic
2
Yes
TENS
TENS at
Yes
maximum
values
Anti-
4
Yes
inflammatory
IACER Srl
Description
Total time 40min
Frequency 90Hz
Impulse width 50µs
Total time 30min
Frequency 1Hz
Impulse width
200µs
Total time 3min
Frequency 150Hz
Impulse width
200µs
Total time 30min
Frequency 120Hz
Impulse width 50µs
Create a square area with the electrodes
over the painful zone. Keep 4cm minimum
distance between the electrodes.
Figure 1 – Electrodes' positioning.