10. Faults
10. Faults
Faults
USB data carrier is full
EN
Error on the USB data carrier
The content of the USB data
carrier cannot be read
Error during transmission of
the measurement
Copy process failed
Measurement could not be
saved
Depot is full
Back pressure too high
Back pressure too low
Input pressure too high
Input pressure too low
Depot pressure too high
The capacity of 500
measurements is reached.
In the presence of strong electromagnetic fields in the frequency range of
260 ... 280 MHz and at 320 MHz, the instrument may be affected under
unfavourable conditions - the display may be disturbed for a short time, i.e.
no numerical values are displayed. This error disappears of its own accord
after a short time. To prevent interference, do not use the instrument in the
vicinity of powerful transmitting devices.
34
Causes
The limit of the USB data carrier's
storage capacity is reached
Access to the USB data carrier
failed.
Error reading the USB data carrier.
Write error on the USB data carrier
Copy process to the USB data
carrier failed.
Write error on the USB data carrier
The intake capacity of the internal
depot is exhausted.
The gas container into which the
measuring gas has been pumped is
under high pressure.
The measured back pressure is too
low.
The minimum pressure is
atmospheric pressure.
The pressure at the inlet is too high.
The pressure at the inlet is too low.
The pressure in the internal gas
depot is too high.
The capacity of 500 measurements
is reached.
WIKA operating instructions analytic instrument, model GA11
Measures
Connect a new USB data carrier or
free up the capacity.
Plug in/pull out the USB data
carrier or use another USB data
carrier.
Plug in/pull out the USB data
carrier or use another USB data
carrier.
Check the USB data carrier via PC,
and format or repair it, if necessary.
Check the USB data carrier and
repeat the process.
Plug in/pull out the USB data
carrier or use another USB data
carrier.
For emptying the depot, (→ see
chapter 5.7 "Emptying depot and
depressurised gas compartment").
Connect a gas container with lower
pressure.
Carry out a pressure compensation
with ambient pressure.
Use external pressure reducer
Use inlet pressure control unit
for gas analytic instruments (e.g.
model GA05).
For emptying the depot, (→ see
chapter 5.7 "Emptying depot and
depressurised gas compartment").
Manually delete the measurements
or activate the option "Automatic
deletion" (→ see chapter
"Settings").