10. Faults
10. Faults
Faults
USB data carrier is full
Error on the USB data carrier
The content of the USB data carrier
cannot be read.
Error during transmission of the
measurement
Copy process failed
Measurement could not be saved.
Depot is full
Back pressure too high
Back pressure too low
Inlet pressure too high
Inlet pressure too low
Depot pressure too high
The capacity of 500 measurements
is reached.
WIKA operating instructions analytic instrument, model GA11
Causes
The limit of the USB data carrier's
storage capacity is reached
Access to the USB data carrier
failed.
Error reading the USB data carrier.
Write error on the USB data carrier
Copy process to the USB data
carrier failed.
Write error on the USB data carrier
The intake capacity of the internal
depot is exhausted.
The gas container into which the
measuring gas has been pumped is
under high pressure.
The measured back pressure is
too low.
The minimum pressure is atmos-
pheric pressure.
The pressure at the inlet is too high. Use external pressure reducer
The pressure at the inlet is too low.
The pressure in internal gas depot
is too high.
The capacity of 500 measurements
is reached.
Measures
Connect a new USB data carrier or
free up the capacity.
Plug in/pull out the USB data carrier
or use another USB data carrier.
Plug in/pull out the USB data carrier
or use another USB data carrier.
Check the USB data carrier via PC,
and format or repair it, if necessary.
Check the USB data carrier and
repeat the process.
Plug in/pull out the USB data carrier
or use another USB data carrier.
Empty the depot (see chapter 6.8
"Emptying depot")
Connect a gas container with lower
pressure.
---
Use inlet pressure control unit for
gas analytic instruments (e.g. model
GA05).
Empty the depot (see chapter 6.8
"Emptying depot")
Manually delete the measurements
or activate the option "Automatic
deletion" (see chapter 6.10.1
"System settings").