8. Maintenance and cleaning
8. Maintenance and cleaning
8.1 Maintenance
Repairs and maintenances, as well as the replacement of parts, must only be carried
out by the manufacturer or manufacturer-approved service partners.
8.2 Cleaning
CAUTION!
Clean the instrument with a moist cloth.
■
Electrical connections must not come into contact with moisture.
■
Wash or clean the dismounted instrument before returning it, in order to
■
protect persons and the environment from exposure to residual media.
Residual media in the dismounted instrument can result in a risk to
■
persons, the environment and equipment.
Take sufficient precautionary measures.
For information on returning the instrument see chapter 11.2 "Return".
8.3 Recalibration
We recommend that the analytic instrument is regularly recalibrated by the
manufacturer. In addition, every factory calibration includes a free-of-charge check of
all system parameters with respect to their compliance with the specification. The basic
settings will be corrected if necessary.
SF₆-Q-Analyser: 2 years
g³-Q-Analyser: 1 year
The time until the next recalibration for each individual sensor can be queried under
"Sensor info".
Settings > System > Sensor info
WIKA operating instructions analytic instrument, model GA11