Emptying The Pump - Medtronic SYNCHROMED II Manuel D'implantation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Filename Date Time
UC200xxxxxx EN
4.6 x 6.0 inches (140 mm x 197 mm)
20-mL empty syringe
Fenestrated drape
Template
Locally supplied:
Syringe containing prescribed fluid (volume of fluid not to exceed the reservoir
volume of the pump)
Cleansing agent
Sterile gloves
Alcohol pads or swabs
2. Prepare the programmer for use. Refer to the Programming Guide packaged with the
pump software for instructions.

Emptying the pump

1. Identify the pump model, reservoir volume, and location of the reservoir fill port and
catheter access port.
2. Confirm that the refill volume of the prescribed fluid does not exceed the reservoir
volume of the pump.
3. Prepare the injection site by cleansing the area.
4. Open the kit and glove packages. Put on the sterile gloves.
5. Place the drape, exposing the pump site.
6. Using sterile procedures, assemble the needle, extension tubing, and empty syringe
(Figure 8).
7. Locate the reservoir fill port septum by placing the template over the pump and aligning
the edges of the template with the edges of the pump.
8. Close the clamp (Figure 8).
9. Gently insert the 22-gauge needle through the template center hole and into the center
of the reservoir fill port septum until the needle touches the needle stop (Figure 8).
#
Caution: Do not use excessive force when accessing the reservoir fill port.
Excessive force can damage the pump, requiring surgical replacement.
1
Included in the Model 856X Refill Kit
M221311A028 Rev A
1
3
3
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Template version: 03-03-2010
8637 2009-10
English 23
2009-10

Publicité

Table des Matières
loading

Produits Connexes pour Medtronic SYNCHROMED II

Ce manuel est également adapté pour:

8637

Table des Matières