Cleaning procedure for Invia Liberty Pump, Docking Station and Charger
1.
Wear suitable protection (clothing, gloves, face mask and goggles) according to local
guidelines.
2.
Apply the disinfectant agent in accordance with instructions from the manufacturer
Pay attention particularly to edges, narrow corners and bottom side.
3.
Leave the disinfectant on. Follow the recommended residence time for the disinfectant
as instructed by the manufacturer
CAUTION
Do not spray disinfectants directly into openings as this may harm electronic
components.
4.
Thoroughly clean the surface, all edges, housing niches, corners, bed holder, brackets,
port covers and bottom side.
5.
Wipe dry or air dry as instructed from the manufacturer
6.
If needed, repeat step 2-5 to ensure proper cleaning.
Dispose contaminated material in accordance with local environmental procedures.
7.
For a detailed Cleaning Instruction, contact your Medela representative.
Manufacturer of the disinfectant agent.
1)
Disposal
Invia Liberty Pump is made from various metals and plastics. Before disposal the rechargeable
battery and electronics must be removed according to instructions. Then Invia Liberty Pump
is no longer operational. Dispose of electronics and plastic components should be handled in
accordance with local guidelines.
Invia Liberty disposables and dressings should be handled and disposed in accordance with
local environmental procedures.
PRECAUTION
Must not be disposed together with household refuse.