– Safety Recommendations and Notes on Maintenance
®
The safety valve should be checked regularly during use twice a month by unscrewing the serrated nut
Build-up of lime in the safety valve can prevent it from functioning. Every 6 months, the following test should also be performed:
fill the Polyclav
to the rim with water, replace and lock the lid and then pressurize. The safety valve must open between 2.5 – 2.8 bar.
®
If the safety valve does not open, unscrew it and clean with vinegar, which will dissolve lime deposits – if necessary, replace the safety
valve.
Should these precautions prove to be ineffective, the unit must be returned to the manufacturer. On safety grounds, only the complete
unit with matching lid and body should be checked – and if necessary – repaired with spare parts from the original manufacturer only.
We recommend that both the interior and exterior of the unit be cleaned regularly to prevent build-up of corrosion products.
We recommend that the periods for retesting registered in the EC Declaration of Conformity be adhered to in your own interest.