Light Signals And Faults - Pepperl+Fuchs Vibracon Mini LVL-A1 Mise En Service

Détecteur de niveau pour liquides
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Light signals and faults

The LED display is on the connection side of the
LVL-A1. The green light always comes on in
service.
Variants AC and DC-PNP with valve plug
Light signal
Red light comes
on
Green light does
not come on
Red light
flashing
Variants AS-i and DC-PNP with M12 x 1 connector
Light signal
Yellow light
comes on
Red light comes
on
AS-i
Red light comes
on
DC-PNP
Green light does
not come on
Red light
flashing (2 Hz)
28
State/Action
Operating mode MAX (overspill protection):
Sensor immersed in fluid.
Operating mode MIN (dry running protection):
Sensor not immersed in fluid.
Error: No power supply.
½
Check plug, cable and power supply.
Error: Overload or short-circuit in load circuit.
½
Rectify the short-circuit.
½
Reduce maximum load current to below
250 mA.
Error: Internal sensor error or sensor
corroded.
½
Replace device.
State/Action
Sensor immersed in fluid.
Error: Address 0 set or communication error.
½
Carry out addressing process.
½
Parameterise slave.
½
Or reduce line length
(< 100 m total length).
Error: Overload or short-circuit in load circuit.
½
Rectify the short-circuit.
½
Reduce maximum load current to below
250 mA.
Error: No power supply.
½
Check plug, cable and power supply.
Error: Internal sensor error or sensor
corroded.
½
Replace device.

Publicité

Table des Matières
loading